Improving a scissor‐action couch for conformal arc radiotherapy and radiosurgery

Abstract

We have developed a method to improve the setup accuracy of a Varian Clinac 6/100 couch for delivering conformal arc therapy using a tertiary micro multileaf collimator (MLC) system. Several immobilization devices have been developed to improve the mechanical stability and isocenter alignment of the couch: turn-knob harnesses, double-track alignment plates, and a drop-in rod that attaches the couch to the concrete floor. These add-on components minimize the intercomponent motion of the couch's scissor elevator, which allows consistent treatment setup. The accuracy of our isocenter couch alignment is an improvement over the above devices, within 1 mm of their accuracy. The couch has been used with over 15 patients and with over 50 modulated conformal arc treatment deliveries at our institution.