Testing the test strips: laboratory performance of fentanyl test strips.

Abstract

BACKGROUND: The overdose crisis driven by synthetic opioids continues to escalate in the USA. We evaluated the efficacy of multiple manufacturing lots of a fentanyl test strip (FTS) to detect fentanyl and fentanyl analogs and assessed cross-reactivity with possible interferences. METHODS: Drug standards were dissolved in water in a laboratory setting and serially diluted. Drug dilutions were tested using five different manufacturing lots of BTNX Rapid Response (20 ng/mL cutoff) lateral flow chromatographic immunoassay strips to assess lot-to-lot variability for FTS sensitivity and cross-reactivity for the analytes of interest. RESULTS: All five manufacturing lots cross-reacted with fentanyl and eleven fentanyl analogs. Diphenhydramine, lidocaine, MDMA, and methamphetamine were found to cause false positives with the strips. There was notable lot-to-lot variability in the sensitivity of the strips for fentanyl, fentanyl analogs, and known interferences. DISCUSSION: FTS remains an important overdose prevention tool, but lot-to-lot variability in performance complicates robust instructions that balance the prevention of false positives and false negatives. Continued lot-to-lot performance assessment is recommended to ensure health education for FTS remains accurate. More sophisticated drug checking technologies and services are needed in the community landscape to augment personal FTS use to facilitate informed consumption and overdose risk mitigation.

Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.

Main Content
For improved accessibility of PDF content, download the file to your device.
Current View